Remove the Positive Control Cassette. Follow the screen instructions to insert the Negative Control Cassette and gently close the drawer as shown for the positive control.
When the QC test is complete, the results will be displayed. Next Step - Run Patient Specimens. The current or default development mode will be shown at the top of the Start Test screen. Sofia will then return to the Start Test screen displaying the newly selected development mode. Enter the Patient ID Use the Barcode Scanner to scan a patient ID barcode from a patient chart or other source, or manually enter the data using the numeric key pad on Sofia.
Enter the Order Number, if applicable Use the Barcode Scanner to scan an order barcode from a patient test requisition or other source or manually enter the data using the numeric key pad on Sofia. In the READ NOW development mode, remember to first manually time the test development of the patient specimen Test Cassette outside Sofia at room temperature for the required period of time as defined in the Package Insert for the specific test before proceeding to the next step.
Insert the Test Cassette into the drawer and gently close the drawer. Upon closing the drawer, Sofia will start automatically and display the progress as shown in the example below. Obtain Patient Specimen Test Results When the test is complete, the results for both the patient specimen test s and for the internal procedural control will be displayed on the Sofia screen.
The results will be automatically printed on the integrated printer if this option has been selected as the default setting. Refer to the Package Insert for each test for details on how to interpret results for specific tests.
In the example above, the internal procedural control was invalid; therefore, the patient specimen test results were invalid, as shown. In the example above, the internal procedural control was valid; the patient specimen was positive for Influenza A and negative for Influenza B.
If the procedural control is "invalid," repeat the test with a new patient sample and a new Cassette. To run another patient specimen, select "Start New Test. The "Switching Off" screen will be displayed, followed by a brief "Please wait Shutdown is complete when the screen goes dark. Next Step - Watch Videos.
The instrument offers an intuitive user experience with a guided touch screen interface, procedural controls, and supervisor overview options. This proven platform in a plate-based format allows for multi-well multiplexing. The QuantStudio Dx instrument provides your clinical laboratory with performance you can trust, security you demand, and flexibility you need in your workflow to meet the needs of your lab today and tomorrow. The well fast block can be used in IVD mode.
Efficient workflows A simplified instrument and assay workflow as well as intuitive software makes the entire test—from processing to reporting—efficient and cost-effective. This industry first feature provides near real-time infectious disease mapping and reporting. Sofia 2 is a small bench top analyzer that uses advanced fluorescence detection with an ultraviolet LED energy source. Sofia is a small bench top analyzer that uses advanced fluorescence detection with an ultraviolet LED energy source.
For use with Sofia 2. The Sofia 2 Lyme Fluorescent Immunoassay FIA uses advanced immunofluorescence-based lateral-flow technology to provide rapid differential detection of For use with Sofia 2 and Sofia.
For use with Sofia Q, Sofia 2, and Sofia. Skip to main content. Explore all our
0コメント